Chapter 6: Maintaining the Nano Enclosure

Product Service 1-800-522-7658

34

8. The table below lists typical alarm conditions based on normal operating

conditions. Typical alarm conditions are set at face velocities of 10 to 30
feet per minute below the normal operating conditions due to supply air
and exhaust air fluctuations, as well as room air cross drafts. Consult your
Safety Officer for proper operating speeds.

Enclosure Operating In-Flow

Speed

Alarm Condition Set Point

Speed

100 ± 10 fpm

70-90 fpm

90 ± 10 fpm

60-80 fpm

80 ± 10 fpm

50-70 fpm

70 ± 10 fpm

50-60 fpm

60 ± 10 fpm

50 fpm