Labconco XPert Nano Enclosures 38876 Series User Manual

Page 50

Advertising
background image

Chapter 7: Accessorizing the Nano Enclosure

Product Service 1-800-522-7658

45

NOTE: The cutout can also be made with a reciprocating saw and a 1/8" diameter

bit. The centerline of the cutout is located 3.3" from the left side and 4.2"
from the bottom. The 1/8" mounting holes are spaced 4.38" apart top to
bottom and centered over the cutout. The cutout dimensions are 2.43"
wide by 4.13" tall.

2.

Canopy Exhaust Connection for Ducting to Outside

Canopy Exhaust connects to the filtered enclosure with integral motorized
impellers from the top so the duct can be routed up and outside the building.
The Canopy Exhaust is available for 6" nominal duct on 2', 3' and 4' sizes and
10" nominal duct on 5' and 6' sizes. The Canopy Exhaust Connection should
be sealed to the top of the enclosure with silicone sealant to prevent leaks. See
Figure 3-3 and Figure B-3.

Order the following:

Part

Description

3924400

2' Canopy Exhaust Connection (6")

3924401

3' Canopy Exhaust Connection (6")

3924402

4' Canopy Exhaust Connection (6")

3924403

5' Canopy Exhaust Connection (10")

3924404

6' Canopy Exhaust Connection (10")

3.

Remote Blowers

Has a 1/4 hp direct drive motor and corrosion-resistant epoxy-coated steel
housing and wheel with blower inlet of 6.00" ID. Outlet dimensions are
4.25" x 7.38" OD.

CFM @ Static Pressure-Inches of H

2

O

S.P.

0.0"

0.125"

0.25"

0.50"

0.75"

0.87"

CFM

595

560

515

420

300

167


Catalog
#

Description

Shipping Wt.

(lbs/kg)

4863500

Remote Blower, 115 V, 60 Hz. 4.4 amps

35/16

4863501

Remote Blower, 115/230 V, 50 Hz, 5.6/2.8 amps

35/16

7053501

Explosion-Proof Remote Blower, 115 V, 60 Hz,
4.4 amps

40/18


Advertising