Chapter 6: Maintaining the Nano Enclosure

Product Service 1-800-522-7658

36

Low Air Alarm

Set Point

(user defined)

Enclosure

Operating

Inflow Speed

Low Calibration

Set Point

(user defined)

High Calibration

Set Point

(user defined)

40 – 50 fpm

60 fpm

40 – 60 fpm

100 – 120 fpm

60 – 70 fpm

80 fpm

50 – 90 fpm

100 – 150 fpm

80 - 90 fpm

100 fpm

50 – 110 fpm

100 – 170 fpm


NOTE: ENTER button stores information

+/- Buttons allow for scrolling.

1. Push the ENTER button on the face of the alarm until the Set up Menu is

displayed.

2. Scroll to SET UP and hit ENTER.

3. The PASSWORD MENU displays (The Password is 0000). Press the

ENTER button repeatedly until the CAL CONFIG MENU is displayed.

4. In the CAL CONFIG MENU set for the following:

CALIBRATION CONFIG MENU

SETTINGS

DISPLAY UNITS

FPM

LOW AIR ALARM

SEE CHART ABOVE – LOW AIR ALARM

LOW AIR CUTOFF

OFF

LOW AIR CUTOFF

SEE CHART ABOVE – LOW AIR ALARM

WARNING AIR ALARM

SEE CHART ABOVE – LOW CALIBRATION SET
POINT (Set 3-5 FPM above low air alarm)

WARNING AIR RESET

3 FPM

HIGH AIR ALARM

OFF

HIGH AIR ALARM

OFF

LOWER AIR SAMPLE FLUCTUATIONS

3%

HIGHER AIR SAMPLE FLUCTUATIONS 3%
LOWER/HIGHER AIR SAMPLE
DIFFERENCE

10 FPM

WARN TO ALARM AIR TIME

10 SECONDS

ALARM TO WARN AIR TIME

3 SECONDS

SHOW AIR FLOW

ON

SHOW TIME LINE OFF = DISPLAYS
BAR GRAPH

OFF

AUDIBLE ALARM

ENABLED

SENSOR DIFFERENCE

2%

5. To complete the CAL CONFIG, be sure to enter “DONE”. You are

returned to the Main Menu.