Chapter 4: Performance Features and Safety Precautions

Product Service 1-800-522-7658

22

17. Increase the face velocity by adjusting the speed control of the motorized

impeller on the XPert Nano Enclosure. Change the filter when the face
velocity falls below acceptable limits established by your Safety Officer.

18. Exhaust ULPA filters are disposed as hazardous waste. The user is

responsible for recording the material and disposing properly. Filters
should be disposed using the disposal bag on XPert enclosures.

19. Ensure only trained operators use the enclosure. New users should review

the User’s Manual and become familiar with the operation of the enclosure.

20. Proper operation of the enclosure depends largely upon the enclosure

location and the operator’s work habits. The enclosure should be located
away from traffic patterns, doors, fans, ventilation registers, fume hoods,
and any other air-handling device that could disrupt its airflow patterns (see
references in Appendix D).

21. For weighing procedures, adjust face velocity between 60 and 100 FPM for

effective operation.

22. Leave the blower on for at least one minute after work in the enclosure has

been completed.

23. If a chemical, powder or particulates are spilled on the work surface, DO

NOT switch off the blower until all traces have been removed.

24. Tag enclosures with appropriate warning, if the enclosure has been disabled

for service.

25. If the blower fails during use, processes should cease and the area should

be vacated and ventilated.