Chapter 5:

Using the Nano Enclosure

Now that the installation of your Nano enclosure is completed, you are ready to
use your Nano enclosure.

Read this chapter to learn about:

Routine Daily Work Procedures

Appropriate ULPA Filter Applications, Suitability and Guidelines

Optional Ionizer

Prohibited Acid Use

Routine Daily Work Procedures

Planning

Arrange for minimal disruptions, such as room traffic or entry into the

room while the enclosure is in use.

Start-up

Turn on exhaust system and light.

Only raise the sash for loading and cleaning.

Check the baffle air slots for obstructions.

Allow the enclosure to operate unobstructed for 1 minute.

Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.

Wear a protective mask if appropriate. Consult your Safety Officer for
additional personal protective equipment recommendations.

Loading Materials and Equipment

Load only the materials required for the procedure. Do not overload the

enclosure.

Do not obstruct the air foil, or rear baffle slots.

Large objects should not be placed close together and should be elevated

above the work surface to permit airflow to sweep under the equipment.

After loading, wait one minute to purge airborne contaminants from the

work area.