Chapter 3: Getting Started

Product Service 1-800-522-7658

11

Connecting to the Exhaust System (Optional)

ATTENTION: The weight of the exhaust ductwork system
must be supported independently of the enclosure
superstructure or damage may occur.

The exhaust system should be installed by a qualified HVAC
contractor
.

The exhaust connection on the 2', 3' and 4' XPert Nano Enclosure have been
designed to accept 6" diameter ductwork as shown in Figure 3-2. The 5' and 6'
models have been designed for 10" diameter ductwork as shown in Figure 3-3.
For the 2', 3' and 4' sizes, the exhaust connection ships uninstalled on the top and
needs to be fastened with the screws provided. See Chapter 7 for ordering Canopy
Exhaust Connection. Canopy Exhaust Connection aids in the removal of
chemicals or applications where a higher degree of powder and particulate removal
is required. Review Chapter 2 for exhaust prerequisites and review Chapter 7 for
ordering blower exhaust equipment.

Consult Labconco Customer Service should you require help sizing your blower
for the exhaust volume and system static pressure loss.

To ensure compatibility, the selected exhaust duct material
should match the enclosure, procedures and chemical
applications.

!

!