Chapter 6: Maintaining the Nano Enclosure

Product Service 1-800-522-7658

40

Motorized Impeller Replacement

The motorized impeller must be replaced as a complete unit. When the motorized
impeller is replaced, the capacitor may also be replaced. See Appendix A for
Replacement Parts Diagram. See Figure 6-5 for an isometric view of the motorized
impeller plenum assembly. The ULPA filter rests on top of the motorized impeller
assembly.

Do NOT contact blower wheel while still in motion.

NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.

1. Wear appropriate personal protective equipment to decontaminate the filtered

enclosure and then unplug from the electrical outlet.

2. Remove filter per the ULPA filter removal procedure outlined in this chapter.

3. Consult the wiring diagram in Appendix C of the manual and disconnect all the

wires of the motorized impeller. Be sure to connect wires on the new motor in
the same way the old motor was wired. If contaminated with chemicals,
neutralize the old motor by submersion in an appropriate solvent. Bag and
dispose of the old motor responsibly.

4. Remove four screws in the motor bracket that hold motor bracket to the

supports with the vibration isolation mounts. Remove the motor and bracket.

ATTENTION: High-speed blower. Never operate impeller with
housing off.

5. Replace the capacitor with a new one of equal voltage and capacity.

6. Reassemble the new motorized impeller by reversing the assembly steps.

!

!