Model Size

2'

3'

4'

5'

6'

Sash Height from Work Surface (inches)

9.44

9.44

9.44

12.00

12.00

Total Open Area with Bypass (sq. ft.)

1.45

2.20

2.95

4.80

5.80

Exhaust Airflow Volume at 90fpm (CFM)

130

200

265

430

520

Alarm Airflow Volume at 60 fpm (CFM)

85

130

175

290

350

Initial Mag. Gauge Pressure Reading (inches H

2

O)

0.20 to .045

0.15 to 0.40

0.10 to 0.35

0.10 to 0.50

0.10 to 0.50

Number of Laskin Nozzles needed at 10 psig

1

1

1

1

1

Exhaust ULPA Filter w/ Gasket Dimensions (inches)

18x18x3.31

30x18x3.31

42x18x3.31

(2) 24x18x3.31

(2) 18x30x3.31

*Motor Horsepower (HP)

0.13

0.13

0.30

0.60

0.60

*Capacitor (MFD, Volts)

10MFD, 115V

(2) 4MFD, 230V

10MFD, 115V

(2) 4MFD, 230V

15MFD, 115V

4MFD, 230V

15MFD, 115V

4MFD, 230V

15MFD, 115V

4MFD, 230V

Fluorescent Lights

F15T8-SP35

F17T8

F25T8

F32T8

F40T8



To determine the actual face velocity at the sash opening, airflow velocity readings
will need to be taken. This should be done across the sash opening of the
enclosure. The “average face velocity” is achieved by taking readings in two rows
across the enclosure with the readings 6" from the ends and evenly spaced every
12"; the first row is 3" down from the upper sash foil and the second row is 3" up
from the work surface.