Filter leak test, Purpose – Labconco XPert Nano Enclosures 38876 Series User Manual

Page 35

Advertising
background image

Chapter 6: Maintaining the Nano Enclosure

Product Service 1-800-522-7658

30























Figure 6-1

Bag-In/Bag-Out Filter Changing Procedure Diagram

& Filter Leak Test Diagram

Filter Leak Test

Purpose

After installing the new ULPA filter, the filter should be leak checked. This test is
performed to determine the integrity of the ULPA filter, the filter housing, and the
filter mounting frames. Leak testing is to be done by a qualified technician with
calibrated equipment. Remove the top perforated exhaust cover by using a Phillips
screwdriver to unfasten the (2) screws used to secure (2) clips. See Figure 6-1. The
filter passes the leak test at .01% or better. Reference Leak Testing and Photometer
scanning from the Institute of Environmental Services (IES-RP-CC001.3)

Apparatus

1. An aerosol photometer ATI model 2D, 2E, 2G or equivalent. Air

Techniques Hamilton Associates Inc., 11403 Cronridge Dr., Owings Mills,
MD 21117

Old Filter

New Filter

Old ULPA Filter Sealed in Old Bag

Filter Gasket Up As Shown

Filter Gasket Up As Shown

Create Two Seals & Cut Here After Sealing

Bag Stub from Old Bag

Clip to Hold Perforated Cover

Perforated Top Cover

Screw (One Each Side)

Speed Control Adjustment

Screw Location

Create Two Seals & Cut Here After Sealing
Old Bag Stub Sealed in Glove of New Bag

New ULPA Filter Shown in New Bag

Advertising