Start up, 2 operating modes – BINDER M 53 User Manual

Page 19

Advertising
background image

M (E2) 10/2014

page 19/71

5.

Start up

After connecting the electrical supply (chap. 4.1), turn on the unit via the main power switch (2).
Warming chambers may release odors in the first few days after commissioning. This is not a quality de-

fect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one

day and in a well-ventilated location.

5.1 Function overview of the MB1 display program controller

Pilot lamp: Ready for operation

EXIT button (to exit a menu point)

AUTOMATIC button (to start a previously entered program)

ENTER button (to confirm a selection)

Navigation buttons (functions are assigned by the menu)

Figure 5: Display program controller MB1

The program controller MB1 controls the temperature inside the drying and heating oven (range: 5 °C

above ambient temperature up to 300 °C).
You can enter the desired set point values in Manual Mode or Program Mode (chap. 5.2) in the display

controller.

W

X

08:43:55 15.12.13

CONFIG

VIEW->

PGM

TEMP

40.0

36.8

°C

HAND

Set point value Actual value

Figure 6: Normal display of the MB1 program controller in Manual mode

5.2 Operating modes

The program controller MB1 operates in 3 modes:

Idle Mode

The controller is not functional, i.e., there is no heating. The fan turns

at a 50% rate.

Manual Mode

(Fixed value operation) (HAND)

The controller operates as a fixed-point control, i.e., a temperature

set-point can be defined, which is then maintained (chap. 8).

Program Mode (AUTO)

An entered temperature program is run (chap. 9).

The program controller MB1 allows programming temperature cycles.
The controller offers 25 program memory positions with 100 program sections each. The total number of

program sections of all programs is limited to 500.