Fault description

Possible cause

Required measures

Miscellaneous
Fan does not turn

Fan speed set to 0%.

Set the fan speed to the desired

value.

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.


15. Technical description

15.1 Factory calibration and adjustment

This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using

standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certi-

fied since December 1996 by TÜV CERT). All test equipment used is subject to the administration of

measurement and test equipment that is also constituent part of the BINDER QM DIN EN ISO 9001 sys-

tems. They are controlled and calibrated in relation to a DKD-Standard at regular intervals.

15.2 Definition of usable volume

The usable volume illustrated below is calculated as follows:

b

c

c

b

a

a

C

B

A

A, B, C = internal dimensions (W, H, D)
a, b, c = wall separation

a = 0.1*A
b = 0.1*B
c = 0.1*C

V

USE

= (A - 2 * a) * (B - 2 * b) * (C - 2 * c)


Figure 15: Determination of the useable volume

The technical data refers to the defined usable volume.

Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them

and thus obtain a homogenous distribution of temperature.