BINDER M 53 User Manual

Page 34

Advertising
background image

M (E2) 10/2014

page 34/71

Program entry as set-point step (example)

W/°C

t/min.

01 02 03 04 05 06 07 08

0

50

100

150

200

250

30

120

180

380

Tolerance

minimum

Tolerance

maximum

On

Off

t/min.

Operation line 1 = air flap

Program table corresponding to the diagram above:

Program

section

Set-point

temp.

Fan

Section

time

Operation

line1

Target

section

No. of

cycles

Min.

tolerance

Max.

tolerance

01

50

100 %

00:30:00

Off

1

0

-1999

+9999

02

50

100 %

00:00:01

Off

1

0

-1999

+9999

03

150

100 %

01:30:00

On

1

0

-5

+5

04

150

100 %

00:00:01

On

1

0

-1999

+9999

05

250

100 %

01:00:00

On

1

0

-2

+2

06

250

100 %

00:00:01

On

1

0

-1999

+9999

07

100

100 %

03:20:00

Off

1

0

-1999

+9999

08

100

100 %

00:00:01

Off

1

0

-1999

+9999

Now enter the values of the above program table into one of the 25 program places of the controller MB1:

For rapid transition phases, do NOT program any tolerance limits in order to allow maximum

heating speed.

PGM

09:17:15 15.12.13