M (E2) 10/2014

page 31/71


You can enter

Program sections into this program table.

Hit the PGM button. An inquiry display appears allowing you to enter or delete individual program sec-

tions:

ZP-Abschnitt Abs. Nr. 5

insert

delete



In this view, new program lines can be entered or deleted:

new

New lines are added below in the table

insert

New lines are added above a previously selected line

delete

Individual lines that have been selected previously are deleted

Create as many lines, i.e. program sections, as desired. As a next step, values can be entered into these

lines. It is possible to add supplementary lines later or to delete individual lines at any time.





To enter values, select the corresponding line via the

arrow keys.
Hit the “ENTER” button. The

program editor ap-

pears.

Enter the individual values of the selected program section.

Program editor Abs.Nr. 6

****.*

Setpoint 1

+100.0

FAN

Operating contacts

Time

00:45:00

Repeat Section

5

Repeat Number

10

Tol.-band min.

-1999.0

Tol.-band max.

+9999.0

Parameter set

1


--

Temperature value at the start of the program section

-- Fan speed in %
-- Operating contact (operation line) 1 = air flap open / closed
-- Duration of the program section
-- No. of start section in case of repeat cycles
-- No. of duplicates in case of repeat cycles
-- Temperature limits (maximum / minimum temperature).

In case

of exceeding: temporary program stop.

-- Pre-selected value (Do NOT change!)

Select the parameters via the arrow keys and confirm by pressing “ENTER”..
Then enter the values via the arrow keys, and confirm the entry by pressing “ENTER”..

Adapt the temperature safety device class 2 (chap. 10.1) or the temperature safety device

class 3.1 (option, chap. 10.2) to the highest temperature set-point value of the program actual-

ly used. Check the safety device for each temperature program and adapt it if necessary.
Set the set-point of temperature safety device class 2 or class 3.1 (option) by about 5 °C to 10

°C above the controller temperature set-point.

Performance after completing the program:
The controller changes to Idle Mode. The heating is inactive; the chamber approximates ambient temper-

ature. The fan turns at a 50% rate.

PGM