BINDER M 53 User Manual

Page 58

Advertising
background image

M (E2) 10/2014

page 58/71

Unit size

53

115

240

400

720

Temperature data (continued)

Recovery time after door

was opened for 30 sec 2)

to 70 °C

min

1

1

1

1

1

to 150 °C

min

3

3

3

3

3

to 300 °C

min

5

5

5

5

5

Air change (air flap open)

at 70 °C

x/h

180

87

57

51

33

at 150 °C

x/h

192

96

60

54

36

at 300 °C

x/h

160

78

54

48

29

Electrical data
IP system of protection acc. to EN 60529

20

20

20

20

20

Nominal voltage (±10 %) 50/60 Hz

V

230 1N~ 230 1N~ 230 1N~ 400 3N~ 400 3N~

Nominal power

kW

1.20

1.60

2.70

3.40

5.00

Energy consumption 3)

at 70 °C

Wh/h

145

230

370

520

570

at 150 °C

Wh/h

300

544

850

1200

1320

at 300 °C

Wh/h

720

1100

1400

2340

2600

Power plug

shock

proof plug

shock

proof plug

shock

proof plug

CEE plug

5 poles

CEE plug

5 poles

Installation category acc. to IEC 61010-1

II

II

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

2

2

Legend:
1) Door without window
2) Up to 98 % of the set value
3) Use this value for sizing air condition systems.
All technical data is specified for unloaded units with standard equipment at an ambient temperature of

+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in ac-

cordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances

of 10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan speed.
All indications are average values, typical for units produced in series. We reserve the right to

change technical specifications at any time.

If the chamber is fully loaded, the specified heating up and cooling down times may vary ac-

cording to the load.