Options, 2 ethernet interface, 3 hepa fresh air filter (option) – BINDER M 53 User Manual

Page 42

Advertising
background image

M (E2) 10/2014

page 42/71


Turn the control knob (3) of the safety device using a coin to its end-stop

(position 10) (chamber protection).

When the set point is reached, turn back the control knob (3) until its trip

point (turn it counter-clockwise).

The trip point is identifiable by the red alarm lamp (3a).

The optimum setting of the safety device is obtained by turning the control

knob clockwise by approx. one scale division, which leads to extinguish the

red alarm lamp (3a).

Figure 10: Setting safety device class 3.1

Check the setting regularly and adjust it following any changes of the set-point.

Function check:
Check the temperature safety device class 3.1 at appropriate intervals for its functionality. It is recom-

mended that the authorized operating personnel should perform such a check, e.g., before starting a

longer work procedure.

11. Options

11.1 Communication software APT-COM™ 3 DataControlSystem (option)

The drying and heating oven is regularly equipped with a serial interface RS 422 that can connect the

BINDER communication software APT-COM™ 3 DataControlSystem. The actual temperature value is

given at adjustable intervals. Programming can be performed graphically via PC. Up to 30 units with RS

422 interface can be cross-linked. For further information, refer to the operating manual of the BINDER

communication software APT-COM™.
Pin allocation of the RS 422 interface:

pin 2: RxD (+)
pin 3: TxD (+)
pin 4: RxD (-)
pin 5: TxD (-)
pin 7: Ground

11.2 Ethernet interface

With this option, the drying and heating oven is equipped with an Ethernet interface that can connect the

BINDER communication software APT-COM™ 3 DataControlSystem. The actual temperature and humid-

ity values are given at adjustable intervals. The MAC Address is indicated below the Ethernet interface.

For further information, please refer to the operating manual of the BINDER communication software

APT-COM™ 3.
With this option, the additional RS422 interface is only used for service purposes. Do NOT connect it to

any network. The interface is labeled accordingly.

11.3 HEPA fresh air filter (option)

With this option, the introduced fresh air is cleaned by means of a high efficiency submicron particulate

air filter type HEPA class H 14 (acc. to DIN EN 1822). Replace the filter insert, if necessary, by removing

the metal cover of the filter at the left side of the unit (Art. No. 6014-0003).