6.3 Menu settings in the “User-settings” menu


User-settings

Instrument data

Contrast

Displ. Power down

Contin. operation

User Level

Safety control. Act + ****.* °C

Safety control. Set. + 0 °C

29

Instrument data

Instrument Name
Enter an individual name of the constant climate chamber.
Address
Enter a controller address (1 to 30) for operation with the communication

software APT-COM™.
All other entries are relevant only for service purposes.

Contrast

(no function)

Displ. power down

Switch off event
Do not change the entry “Wait. Period”.
Waiting period
You can enter a delay time after which the display, following manual activa-

tion, will automatically be turned off. This happens when the moment is out-

side the operation time defined in menu ”Contin. operation”.

Contin. operation

Enter an operation time to determine the period of display activity. Outside

the defined time, the display is automatically turned off. Pressing down any

key will reactivate the display. After the time set in menu “Displ. power down“,

the display will turn off again when the actual time is not within the operation

time fixed in menu “Cont. operation“.

User Level

Toggle here to the display menu “User Level” (chap. 6.4) by entering a pass-

word. Factory default setting for this password is +00001. You can change

the password (“user code”) in the menu “User Level”.

Safety control.Set

The safety controller is not used with the actual controller version. The dis-

plays are without function.

Safety control.Act

The safety controller is not used with the actual controller version. The dis-

plays are without function.