Casella CEL CEL-414 User Manual

Page 10

Advertising
background image

CEL-328/3, or to an optional external filter
set connected to the sound level meter
via the auxiliary socket, so that frequency
analysis can be performed.

Integrated RMS detectors are employed
that provide a full 120 dB measurement
range in three overlapping 60 dB dy-
anamic ranges (100 dB ranges on the
CEL-493/3). The standard Fast, Slow,
Impulse, and Peak time weightings can
be selected. The RMS output signal
passes through an analog to digital con-
verter to the built in low power microproc-
essor which drives a custom designed
liquid crystal display with both analog and
digital indication.

The digital display can be switched to
show the sound pressure level, or any of
the calculated parameters with a resolu-
tion of 0.1 dB. The analog display is a bar
graph that always gives an indication of
the instantaneous sound pressure level

with 1 dB resolution. In order to display
the whole 60 dB (or 100 dB) dynamic
range of the instrument on a 26 segment
bar, the analog display is auto-ranged in
20 dB steps with a reference datum.

Control of the display and calculation func-
tions is undertaken by the microproces-
sor. Not only is a continuous display of
the sound pressure level provided at all
times, but the detector output is also
sampled 320 times per second, and the
samples integrated by the microproces-
sor (not CEL-328/3E or CEL-414/3).

Simultaneous calculations are made to de-
termine each of the parameters available
from the instrument. The results are
stored in registers, and may be shown on
the digital display.

Special control sequences are avoided by
using simple push button controls to se-
lect any required parameter by stepping

Figure 1: Block diagram

920001

CEL-192 Microphone

30 dB

Gain

A-Wtg

Lin. or
C-Wtg

Freq
Weighting

F

Amplifier

Cal

External Power Supply

Power

Supply

Batteries

Display

A / D

Converter

Micro-

processor

Low
Voltages

Internal

Controls

Battery
Condition

RAM (not

414 or

328/3C)

RMS

Detector

EPROM

Address Bus

Data Bus

Log DC Out

80

P

I

S

C-Wtg

CEL-225/3 Preamplifier

200 V Polarisation Voltage

Interface

(only B, D &

493/3)

To Ext Filter: AC Out

From Ext Filter
(not 266/328)

Type 1

10 dB Attenuator Pad
(493/3 only)

Type 2

10 Octave

Filters

CEL-230 Microphon/
Preamplifier Unit

50

20
Range

(only 266/328)

Lin, A-, or
C-Wtd AC Out

30 dB

Gain

Page 6 - General CEL Sound Level Meters Handbook