Casella CEL CEL-414 User Manual

Page 27

Advertising
background image

connecting link has been fitted as de-
scribed in Section 10.2.

1. 1SM measurement is selected auto-

matically at the end of the self test
instead of SPL when the link is in
place, otherwise use the FUNCTION
buttons to step the arrow marker
round the display to “1SM”.

The digital display will show the 1SM
profile level (updated every second),
while the analog bar continues to
show the sound pressure level.

6.4Duration of Measure-

ment Period - DUR

1. Use the FUNCTION buttons to step

the arrow until it points to “DUR”.

The time elapsed since the instrument
last experienced a general reset or
function reset will be shown on the
digital display with an accuracy
better than 5 s in 24 hours.

The analog display continues to show
the instantaneous sound pressure
level.

6.5Measurement of Maxi-

mum Value - L

max

1. Use the FUNCTION buttons to step

the arrow until it points to “MAX”.

The highest sound pressure level
measured during the period since
the L

max

register was last reset will

be shown on the digital display with
a resolution of 0.1 dB.

This level is updated each time a new
maximum is reached, and does not
decay with time as it is stored
digitally.

2. Reset the L

max

register either by

means of a general instrument reset
(see Section 6.1, step 7 above), or
by pressing the RESET button while
“MAX” is selected for display
(Section 6.1, step 8).

Refer to Sections 3.2 and 3.3 for
further information on the reset
possibilities.

The analog bar graph always shows
the instantaneous sound pressure
level.

6.6Measurement of

Equivalent Continuous
Level - L

eq

(not

CEL-328/3E or CEL-414/3C)

Measurement of broadband L

eq

is not

possible with CEL-328/3E, which can
measure L

eq

only in octave bands (see

section 6.10).

1. Use the FUNCTION buttons to step

the arrow until it points to “LEQ”.

The choice of a time weighting will
probably be governed by the need to
obtain a maximum L

max

value during

the same period, so fast or slow
time weightings are normally
selected.

However a value for the impulse
weighted average value L

AIm

can be

obtained when impulse weighting is
selected.

The equivalent continuous level
measured during the period since
the L

eq

register was last reset will be

shown on the digital display with a
resolution of 0.1 dB, while the
instan- taneous sound pressure level
is indicated by the analog display.

All

Except

328/3E

414/3

General CEL Sound Level Meters Handbook - Page 23