in the CEL-438, CEL-338 and CEL-238A
series, or to some other digital device for
data logging or processing, and for con-
trol purposes.

With the exception of the CEL-414/3C
which has no integration functions, these
instruments have the following features:

A

Standard European version of the
instrument measuring SPL, L

eq

,

L

Tm3

, L

Tm5

L

max

, SEL, (and 1SM),

with A-weighted and Linear
frequency response,

B

Standard European version (as A)
that contains a CEL low power serial
interface,

C

Standard American version of the
instrument measuring SPL, L

eq

,

L

OSHA

, L

DOD

, L

max

SEL, (and 1SM),

with A- and C-weighted frequency
response,

D

Standard American version (as C)
that contains a CEL low power serial
interface,

E

(CEL-328/3 only) Non-integrating
version based on the CEL-328/3C,
without L

OSHA

, L

DOD

, or broadband

L

eq

.

All instruments and versions have a simi-
lar appearance, with minor differences
only in the legends on the frequency
weighting switch and on the bezel sur-
rounding the main display. The best way
to identify an instrument version is to re-
fer to the suffix letter in the instrument
serial number on the rear cover, or to
note the version letter (A, C or E in upper
case, b or d in lower case) displayed to-
gether with the type number during the
instrument’s start up sequence.

All versions will accept the CEL-3025 Vi-
bration Attachment connected via the

Table 1: Functions available with various time weighting selections

Version

Time Weighting

A, B & C & D

E

414/3

Fast Slow ImpulsePeak

493/3

Sound Pressure Level (SPL)

X

X

X

X

X

X

X

Broad band Energy
Equivalent Level (L

eq

)

X

X

X

X

1

X

2

Octave band Energy
Equivalent Level (L

eq

)

X

X

X

X

X

1

Maximum Value (L

max

)

X

X

X

X

X

X

X

X

Measurement Duration (DUR)

X

X

X

X

X

X

X

X

Sound Exposure Level (SEL

3

)

X

X

X

X

Taktmaximalpegel 3s (L

Tm3

)

X

X

X

Taktmaximalpegel 5s (L

Tm5

)

X

X

X

L

OSHA

X

X

L

DOD

X

X

1SM

4

X

X

X

X

X

Note

1

: Slow L

eq

does not fully satisfy the requirements of IEC 804.

Note

2

: Impulse time weighted L

eq

is used only when complying with VDI 2058.

Note

3

: Not available with Types CEL-266/3, CEL-328/3 and CEL-414/3.

Note

4

: 1SM is a special application required by IEC 651 para 7.8 which calls for a one

second maximum profile to be adopted for the SPL function. Measurements in this mode
can be made when the link is connected according to Section 6.3.