Digital display:

3

1

/

2

digit, 7 segment with 0.1 dB

resolution.

Analog display:

26 segment bar, with 1 dB resolution.

Functions indicated:

SPL

Sound Pressure Level,

L

eq

Equivalent Continuous
Level (not CEL-414/3),

L

Tm3

Maximum Value
averaged over 3 sec
periods, (A, B &
CEL-493/3),

L

Tm5

Maximum Value
averaged over 5 sec
periods, (A, B &
CEL-493/3),

L

OSHA

Average A-weighted
slow level with a
preset threshold and q
= 5 (C & D versions
but not CEL-414/3),

L

DOD

Average A-weighted
slow level with a
preset threshold and q
= 4 (C & D versions
but not CEL-414/3),

L

max

Maximum Level,

DUR

Duration over which
measurements taken,

SEL

Sound Exposure Level
(not CEL-266/3,
CEL-328/3 or
CEL-414/3)

1SM

One Second Maximum
when special link is
connected according
to Section 10.2 (not
CEL-266/3 or
CEL-328/3),

31-16k

Octave band centred
on the indicated

frequency (only
CEL-266/3 and
CEL-328/3)

BAT

Battery voltage.

Parameters displayed:

dB,
Battery voltage,
Hours & minutes,
Low battery level,
Overload,
Frequency weighting in use,
Filter in use,
PAUSE button operated,
Error messages.

Impulse capability:

63 dB peak to RMS, 1 ms of 4 kHz.

Start up stabiliaztion time:

90 seconds.

Overload detection:

Pre- and post-weighting detectors,
with memory.

Analog outputs (via auxiliary
socket):

AC (CEL-493/3 only):

Conditioned signal,
typically 8.0 V RMS for
FSD.
Recommended imped-
ance of any connected
device (including
interconnecting cable)
>10 kΩ and <50 pF.
Consult CEL about
cables for use with
equipment not
manufactured by CEL.