practice associated with the Health and
Safety at Work Act, and the Control of
Pollution Act.

Many British Standards have international
equivalents (for example ISO 1996 Noise
Nuisance, ISO 1999 Noise Deafness,
etc), and most countries have adopted
their own national standards along similar
lines. Refer to the relevant standard for
detailed information about any particular
measurement procedure.

The measurement parameters available
on these sound level meters are as fol-
lows:

L

eq

Equivalent Continuous
Level (not CEL-414/3
and only octave band
L

eq

with CEL-328/3E),

SPL

Sound Pressure Level,

L

max

Maximum Level,

DUR

Duration over which
measurements taken,

L

Tm3

Maximum A-weighted
Average values taken
from a series of 3
second periods (A, B &
CEL-493/3),

L

Tm5

Maximum A-weighted
Average values taken
from a series of 5
second periods (A, B &
CEL-493/3),

L

OSHA

Averaged level with a
preset threshold and
q = 5 (C & D versions
but not CEL-414/3),

L

DOD

Averaged level with a
preset threshold and
q = 4 (C & D versions
but not CEL-414/3),

SEL

Sound Exposure Level
(not CEL-266/3,
CEL-328/3 or
CEL-414/3),

1SM

One Second Maximum
(not CEL-266/3,
CEL-328/3 or
CEL-414/3), requires a
link to be connected
(see Section 10.2), and
the instrument to be
set as for SPL
measurement,

31-16k

Octave bands centred
on indicated frequency
(CEL-266/3 and
CEL-328/3 only).

6.1General Procedures

Operation is commenced as follows.

1. Insert batteries and connect the

microphone and preamplifier
according to Chapter 4.

2. Check the battery condition by sliding

the ON-BAT-OFF switch to “Bat”.

The instrument takes approximately
10 s to perform its self test routine,
after which the battery voltage will
be indicated on the main digital
display. Batteries in good condition
should give between 5.8 and 6.2 V.

3

If the reading is close to the 3.2 V
level at which the“LOW BATT”
message is displayed, load fresh
batteries. (Refer to Section 4.1.)

4. If starting from cold, wait 90 seconds

for all measurment and integration
circuits to stabilize at their specified
accuracy.

5. Perform a field accuracy check, i.e.

calibrate the instrument acoustically
following the procedure given in
Chapter 5.