Casella CEL CEL-414 User Manual

Page 33

Advertising
background image

The C5681/1 (1 m) cable supplied is also
used to connect the arrangement to a
CEL-284/2 Calibrator for electrical calibra-
tion at 100 mV. Refer to Figure 12 for ca-
bles to be used with the earlier type
CEL-177 Calibrators.

Vibration is detected by the accelerome-
ter which emits a voltage signal propor-
tional to the acceleration level. The
voltage is passed via the integrator to the
sound level meter where it is measured
and displayed as a decibel level.

One or two stages of integration may be
selected, so the meter may display dis-
placement, velocity, or acceleration levels
that can be easily converted to appropri-
ate units using the slide rule provided.

7.3Recording With CEL-

238A, CEL-338, & CEL-
438 Digital Recorders &
Analysers (B, D &
CEL-493/3)

These instruments contain a CEL low
power interface and can be connected to
the CEL-238A Secondary Processor, CEL-
338 Graphics Printer & Logger, and CEL-

438 Noise Level Analyser which store
data digitally and have their own micro-
processors to manage all the control and
calculation procedures they can perform.

A standard C4530/2 (2 m) Cable connects
the CEL digital recorder/ analyser to the
sound level meter auxiliary socket as
shown in Figure 13.

The auxiliary sockets of CEL filter sets
have similar pin identities, so that the
same connecting cable can also be used
with them.

920006

CEL-225/3 or CEL-225

CEL Sound Level Meter

Calibrator
(With Microdot)

Earlier CEL-177

CEL-284/2

Calibrator

C5681/1 Cable
(Supplied)

Calibrator
(With Jack Plug)

Later CEL-177

Preamplifiers

C3849/1 Jack
Plug Cable

CEL-189

Integrator

C3150/1.5 Microdot
Cable (Supplied)

Figure 12: Connections for

electrical calibration

920009

C4530/2

CEL-238A

CEL-338

CEL-438

Figure 13: Connection with

CEL-238A, CEL-338 or CEL-438

Data Loggers

General CEL Sound Level Meters Handbook - Page 29

Advertising
This manual is related to the following products: