Table 4: Measurement range of CEL-493/3

Range

Freq

Display

Primary

Primary

Pulse

Setting

Wtg

Range

Linearity

Linearity

Range

Range*

Range

Peak

IEC 651

IEC 804

Value
IEC 804

Lo

A

10-111.1 dB

27-91 dB

27-110 dB

86 dB

Linear

10-111.1 dB

45-91 dB

45-110 dB

68 dB

Med

A

40-141.1 dB

43-121 dB

43-140 dB

100 dB

Linear

40-141.1 dB

50-121 dB

50-140 dB

93 dB

Hi

A

50-151.1 dB

53-131 dB

53-150 dB

100 dB

Linear

50-151.1 dB

53-131 dB

53-150 dB

100 dB

Note*: The lower reading is quoted at +10 dB on noise floor.

The upper reading allows crest factor 10.

Table 3: Measurement ranges - All instruments except CEL-493/3)

Range

Freq

Display

Primary

Primary

Primary

Pulse

Setting

Wtg

Range

Linearity

Linearity-

Linearity

Range

Range*

Range*

Range

Peak

IEC 651

IEC 651

IEC 804

Value

(Type 1)

(Type 2)

IEC 804

20

A

20-81,1 dB

27-61 dB

27-67 dB

27-81 dB

57 dB

Linear

20-81.1 dB

45-61 dB

45-67 dB

45-81 dB

39 dB

C

20-81.1 dB

31-61 dB

31-67 dB

31-81 dB

54 dB

50

A

50-111.1 dB

51-91 dB

51-97 dB

51-111 dB

63 dB

Linear

50-111.1 dB

51-91 dB

51-97 dB

51-111 dB

63 dB

C

50-111.1 dB

51-91 dB

51-97 dB

51-111 dB

63 dB

80

A

80-141.1 dB

81-121 dB

81-127 dB

81-141 dB

63 dB

Linear

80-141.1 dB

81-121 dB

81-127 dB

81-141 dB

63 dB

C

80-141.1 dB

81-121 dB

81-127 dB

81-141 dB

63 dB

Note*: The lower reading is quoted at +10 dB on noise floor.

The upper reading allows crest factor 10 for Type 1 instruments, and crest factor 5 for
Type 2 instruments.

General CEL Sound Level Meters Handbook - Page 45