14. If the recording levels are lower than

90 dB, record the 114 dBsignal from
the calibrator as described in step
11. Then without touching the
record level on the tape recorder,
move the 80-50-20 Switch on the
sound level meter (not CEL-493/3) to
“50" for signals in the range 50 to
110 dB, or ”20" for signals between
20 and 80 dB.

When these signals are replayed, 30

or 60 dB respectively must be sub-

tracted from the values indicated by
the sound level meter to obtain the
true level.

15. Record about 30 seconds of the

calibration signal.

16. Switch the calibrator OFF and remove

it from the microphone.

17. Adjust the A-L-FA-FL/A-C-FA-FL

switch on the sound level meter
(and the frequency of the filter set)
to the settings required for the
particular recording task.

18. Proceed with recording the noise or

vibration signal.

If the recorded material is to be used
for frequency analysis, it is recom-
mended that each tape be calibrated
individually by using a suitable
generator to produce signals at the
required (centre) frequencies.
Octave or third-octave filtered
output signals from a CEL-213
Random Noise Generator may be
used for this calibration.

When replayed and analysed, these
signals will give an indication of the
record/replay frequency character-
istics of the tape.

7.4.2 Replay From a Tape

Recorder

These guide lines are recommended for
replaying sound or vibration signals with
a CEL sound level meter.

1. Prepare the sound level meter (and

filter set if connected) for use as
detailed in Chapter 4.

2. Using a suitable acoustic calibrator,

perform a field accuracy check on
the sound level meter (and filter set)
as described in Chapter 5.

3. Connect the sound level meter with

the tape recorder as shown in Figure
17 for replay.

4. Set the sound level meter controls as

follows:

P-I-F-S

F (Fast),

A-L-FA-FL or

A-C-FA-FL

L (Linear), C (C-weighted)
or A (A-weighted) if no
filter is to be used,

UHER 4000

CEL-216

C3526/2

920011

CEL-316

Figure 17: Connection with a tape

recorder (Uher 4000) for replay