Table 1: Available Functions

Function

Name

Example

Parameters

Input

Channels

Result

none

none()

None

Input channel

No additional

processing is

performed,

output value

equals input

value

minhold

minhold()

None

1. Input Channel Holds most

negative value

until reset

maxhold

maxhold()

None

1.Input Channel Holds most

positive value

until reset

rollingavg

rollingavg(60)

1.Averaging

period in

seconds

1.Input Channel Calculates

rolling average

over given

interval

vectoravg

vectoravg(60)

1.Averaging

period in

seconds

1.Input Channel Calculates

vector average

over given

interval

degreeday

degreeday(30,gain) 1.Limit value

2.gain -

accumulate

when value is

above limit

loss -

accumulate

when value is

below limit

1.Input Channel Calculate solar

gain (loss)

based on

percentage of

time (over 24

hours) that the

value exceeds

(is below) limit

windto

windto()

None

1.Input Channel Convert input

value of wind

from direction

to output value

of wind to

direction

crosswind

crosswind()

None

1.Wind speed

channel

2.Wind

direction

channel

3.Bearing

channel

Calculate cross

wind

component for

wind speed

using direction

and bearing

values

vapour

vapour()

None

1.Air

temperature

2.Humidity

Calculate the

saturated

vapour

pressure for

the air

temperature

and humidity

values