11.

POWER SUPPLIES

11.1 Solar Power

The Sensus Logger incorporates a solar

power regulator as standard. This

regulator is designed to operate in

conjunction with a nominal 12 V (30 W)

panel with a short circuit current of no

more than 2 A.

The solar panel should be wired

to the connector identified in Figure 25.

The pin assignments for the

solar panel connector are as follows.

Pin

Assignment

1

Solar panel negative terminal,

2

Solar panel positive terminal.

11.2 Battery

The Sensus Logger incorporates a 12 V

lead-acid battery charger as standard.

This charger is designed to operate in

conjunction with a nominal 12 V lead-acid

battery and supplies a maximum charge

current of 0.5 A.

The lead-acid battery should be

wired to the connector identified in

Figure 26.

The pin assignments for the 12

V lead-acid battery connector are as

follows.

Pin

Assignment

1

Battery negative terminal,

2

Battery positive terminal.