Alarm Output Channels:

4 alarm output channels

2 available from the main connector panel 2 via expansion port
Primary channels (main connector):
These have open-collector outputs, suitable to drive relays on external

equipment.
Secondary channels (expansion port):
These are 5 V logic outputs, which require additional buffering

between them and any external equipment.

Alarm Channels:

16 alarm channels, each of which is capable of activating any of the

‘alarm output channels’.
Each alarm channel can be programmed to act individually or in

combination with other alarms to cater for complex alarm conditions.
Each channel includes:
Identity (Free text, used when notifying external equipment of alarm

conditions) Sounder (used to notify alarm condition) Test mode

(Identifies the type of comparison to make)
EQ

data equal to limit value

NEQ

data not equal to limit value

GT

data greater than limit value

LT

data less than limit value

LT or GT

data less than first limit or greater than

second limit value

GT and LT

data greater than first limit and less than

second limit

Action (Identifies how alarms interact)
NONE

alarm is independent from any others

OR

alarm result is “or’d” with result of

next alarm

AND

alarm result is “and’ed” with result of

next alarm

Fast mode alarm can trigger a change in both the channel scan and

data logging rates, in the presence of an alarm condition.
State delays where each change of ‘alarm state’ can have a delay

allocated to it, thereby no action is taken until the alarm state has

been steady for the specified time.
Data values that can be assigned for alarm-on and alarm-off; these

get transferred to a specified data channel when the alarm’s state

changes.