All of these parameters may be changed using commands built into the

logger software interface.

6.4

RS 485 Command Alternate

The Sensus Logger also has an RS 485

interface accessible from the connector array

(Figure 19). The signals are available from a

2-way spring terminal block.

This interface can be used as an

alternate to the primary command interface

(RS232) where longer distances exist.

The pin connections for the RS 485

alternate command interface connector are as

follows.

Pin

Assignment

1

A (-)

2

B (+)

The Sensus Logger is shipped from Casella CEL with this instrument

interface set to the following parameters.

Default parameters - RS485 alternate command interface

Baud rate

9600

Data bits

8

Parity

None

Stop bits

1

Handshake

None

All of these parameters may be changed using commands built into the

logger software interface.

6.5

Modems

The Sensus Logger is pre-configured with a generic set of command/

response strings suitable for most Hayes-compatible modems. If the modem

used differs from this standard, these strings will need to be altered to match

the new configuration.

In addition, all command/response time-out values are also

configurable for those instances where the default values are insufficient for

the equipment used.

Communications

01081

Pin 1
Pin 2

Figure 19: RS 485 interface

SENSUS Digital Data Logger -
Users Handbook