For data type channels, this is the channel number of another data

channel whose data values will be further processed by this channel. The

allowable range is 1 to the maximum number of data channels.

For parameter type channels, this is the number of the parameter

whose value is to be further processed by the channel. The allowable range

is 1 to 100.

3.2.2 Single-ended or Differential

This option applies only to analog type channels and determines whether the

associated analog channel will be read as a single-ended voltage input or as a

differential input. It should be set to match how the sensor is connected.

3.2.3 Gain Setting

Again only for analog type channels, it sets the gain to be applied to the input

analog signal before the logger converts it. The allowable options are shown

below.

Analog Gain Option

Action

x1

Leave analog signal untouched,

full scale ±2.5 V.

x10

Multiply analog signal by a factor

of 10, full scale ±250 mV.

x100

Multiply analog signal by a factor

of 100, full scale ±25 mV.

3.2.4 Report Count or Frequency

This applies only to counter type channels and determines whether the

channel result will represent the pulse count for the measurement period or

the equivalent frequency value.

3.2.5 Counter Reset Mode

Again only for counter type channels, it sets when the count value is to be

reset. The allowable options are shown below.

Counter Reset Option Action

Reset on scan

Reset the counter every time it is read in

the scan event.

Reset on log

Reset the counter every time the log

event occurs.

Reset at midnight

Reset the counter only at midnight

(for daily accumulated values).

3.2.6 Select Instrument Type

This option is only for instrument type channels and specifies the type of

instrument attached to the logger. It configures how the logger will talk to

the instrument and which data values it is to extract.

3.2.7 Set Data Value

Purely for passive type data channels which do not actively gather data from

other sources but merely presents data for other channels to read. This

option allows the data value held to be updated.

Commands

SENSUS Digital Data Logger -
Users Handbook