Load monitoring (lm) – d&b D80 User Manual

Page 58

Advertising
background image

3. As a next step, execute a

System calibration.

Þ The calibration process identifies the actual load

impedances for each channel. The result will be stored as

a reference and used to calculate the upper and lower

limits for the tolerance band.
To verify the correct connection of cabinets and amplifiers,

the calibration result can be compared with the typical

impedance values for d&b loudspeakers as listed in the

Þ Appendix: " Typical impedance (Z) values" on page

76.

4. Run

System check

Þ Executing a System check after the event will repeat the

measurement and display all values that are outside of the

tolerance band, indicating a possible damage to the

system components.
When a sound system is repeatedly used in the same

configuration for which a calibration file exists, the System

check procedure can also be used before the show to

verify the system's correct setup.
A System check will only provide valid impedance values,

if the amplifier has been calibrated before the check with

the load connected or if a valid calibration file has been

loaded using R1.

10.7.2. Load monitoring (LM)
Features
The d&b Load monitoring function is related to the System check

function and serves to identify a possible loudspeaker malfunction.

Function
The calibration process, performed from the System check menu

with the completely set up system, determines the impedance for

each channel and calculates the related upper and lower

impedance limits.

While the system is operating, d&b Load monitoring continuously

monitors the load impedance separately for both frequencies. It

detects any changes in loudspeaker impedance and reports an

error if the limits are exceeded. For this purpose, Load monitoring

uses inaudible pilot signals which are faded in for approximately 2

seconds in user defined intervals.

NOTICE!

The resolution of Load monitoring regarding failures of single

components depends on the type and the number of loudspeakers

connected to each channel.
The maximum number of cabinets which can be operated in

parallel by one amplifier channel in order to be able to detect a

failure of a loudspeaker component is listed in the Þ Appendix:

" Max. number of cabinets operated in parallel" on page 78.
Load monitoring does not work if:

the amplifier is switched off or switched to standby mode.

the respective channel is muted.

the pilot signal level is too low.

d&b D80 Manual 1.9 en

58

Advertising
This manual is related to the following products:

10D, 30D, D20