INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

98

12.2. INTRINSICALLY SAFE MOUNTING

The mounting of PR-01-S-FM must be in accordance with the National Electrical code (ANSI/NFPA70).
For installation guidance of the PR-01-S-FM, see also ANSI/NFPA 12.6 “Installation of Intrinsically safe
instrument systems in Class I Hazardous Locations”. Choose the mounting location of the sensor, barrier
unit and indicating transmitter so that they are protected from sudden impact and friction. If any of the
system parts is affected by a sudden impact, power it off immediately and have it checked by trained
K-Patents service personnel before it's used again.

The electrical connections for PR-01-S-AX are described in Figure 12.21, for PR-01-S-FM in Figure 12.22
and for PR-01-CS in Figure 12.23 for following two alternatives:

– Connections by instrumentation cable

– Connections by armored instrumentation cable

Cables

Following cable is included in a standard delivery:

– 10 m (33 ft) cable Indicating transmitter - Barrier power unit.

Part PR-8040-010.
The maximum cable length is 100 m (330 ft).

– Cable connector to Sensor with screw terminals.

Part PR-8031 with instrumentation cable glands.
Part PR-8032 with armored glands.

The intrinsically safe cable can be connected to the cable connector as described in Figures 12.21 and
12.22 and 12.23.

Options:

– Cable between Barrier unit and sensor, Part PR-8060, armored instrumentation cable with connector.

– Cable between Barrier unit and sensor, Part PR-8050, shielded instrumentation cable with connector.



For cables between Barrier power unit and Barriers, see Section 12.3.