K-Patents PR-01-S User Manual

Page 64

Advertising
background image

INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

62

d

SUPPORT RING

PRISM
KALREZ GASKET
SENSOR BODY

Figure 6.71

Mounting of Kalrez gaskets.

3. Always use new gaskets. Clean the gaskets with cleansing spray. Mount the Kalrez gaskets and put

the prism in place. Slide the prism lightly in the groove in both directions (± 1 mm). This helps the
gaskets to settle in the right position.

4. Put prism holders and screws in place. The teflon or compression pads should be carefully fitted

between the prism and the prism holders. No part of the compression pads should be between the
prism holder and the refractometer body. Tighten the screws alternatingly in small steps. Trim excess
teflon with a razor blade.

5. Tighten the screws with torque wrench to 10 Newton meters. Note. For the special wetted parts

materials, such as Hastelloy C 276, Titanium and SAF 2205, the torque is 5 Newton meters.

6. Inspect the gasket contact surfaces through the prism. They should form regular concentric rings

around both holes (Figure 6.72). The rings should have the same size.

7. If the contact surfaces do not look good loosen the screws and start again from the beginning.

PRISM

GASKETS

Figure 6.72

Kalrez gaskets seen through the prism

Advertising