8.5. RECOMMENDED WASH PRESSURES AND TIMES

To select a recommended wash pressure use the following table:

Nozzle type

Integral nozzles

Wash medium Wash time

seconds

Minimum above
process pressure

Maximum above
process pressure

- HPS

steam

2-3

2 bar (30 psi)

6 bar (85 psi)

- HPN

pressure water

10-15

30 bar (450 psi)

70 bar (1000 psi)

HIMP-2

steam

3-5

2 bar (60 psi)

6 bar (85 psi)

HIMP-3

water

10-15

30 bar (450 psi)

70 bar (1000 psi)

Flow cell nozzles

PR-3356 / WP

pressure water

10-15

30 bar (450 psi)

70 bar (1000 psi)

PR-3355 / SN

steam

3

2 bar (30 psi)

4 bar (70 psi)

PR-3354 / WN

water

10

2 bar (30 psi)

4 bar (70 psi)

Table 8.50

Recommended wash pressures and times.

Note. Steam wash:

Do not use longer wash times than is recommended in the Table 8.50. In case of
coating adjust the wash interval.