K-Patents PR-01-S User Manual

Page 65

Advertising
background image

INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

63

Prism mounting with teflon gaskets

Teflon gaskets are mounted the same way as Kalrez gaskets. Do not handle the teflon gaskets with metal
tools like pliers or tweezers.

False leakage

Sometimes process liquid can be seen behind the prism, Figure 6.73. This is normal and does not indicate
that the prism gaskets are leaking. Only the round areas inside the prism gasket rings are sealed from the
process liquid. Thus leakage is indicated only if process liquid is seen in the area inside the rings.

FALSE

LEAKAGE

Figure 6.73

False leakage.

6.8. SENSOR DISASSEMBLY

Cover

Open the clamp using a 11 mm wrench. Lift cover carefully, the O-ring seal causes some friction.

Image digitizer card

Remove the Image Digitizer Card (see Figure 6.52). Note that all connectors have latches. The color code
for the terminal strip is given by Figure 6.52. Do not touch any other screw than A in Figure 6.80 and B
in Figure 6.81.

Image detector module

Remove Screw A, Figure 6.80. Then the holder can be removed. Gently pull the module out. Be careful
not to pull the Image detector card.
Pull from the back of the module body. At assembly note the
aligning pin. Do not touch the screws G or F, because they fix the Image detector card with the CCD-
element for receiving optical image correctly from the optics inside of the module.

Light source module

Remove Screw B, Figure 6.81. Then the module can be pulled out. (The hole C should always be empty).
The emitter assembly is locked by screw D and can be removed from the module.

Temperature sensor

Use fingers to turn the Temperature Sensor (Fig. 6.81) counterclockwise.

Advertising