INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

29

3

2

2

3

7

5

SD

RD

SGND

SGND DAT- DAT+

Receive data
Transmit data
Signal ground

INDICATING TRANSMITTER

D25 D9

DCE

Figure 3.70

RS-232 connection to DCE-type equipment (e.g., modem).

2

3

4

7

8

1

3

2

5

8

20 4

7

5

6

6

Transmit data
Receive data
Signal ground

INDICATING TRANSMITTER

Request to send
Clear to send
Data set ready
Data carrier detect
Data terminal ready

SD

RD

SGND

SGND DAT- DAT+

D25 D9

DTE

Figure 3.71

RS-232 connection to DTE-type equipment (e.g., computer).

The character structure conforms to the ISO 1177 standard. It is compatible with the 'RS-232' interfaces of
personal computers.

The character parameters are configuration selectable:

speed 1200 to 9600 bits/s
parity odd, even or none
stop bits 1 or 2
flow control hardware, XON / XOFF or none.

The set of measurement results is output in a fixed-length record. The record consists of variable-length
numeric fields at fixed locations in the record. The gaps are filled with ASCII spaces (code 20 hex).

The record format is (program version 4.0 and newer):

Column Field

Format

Value

1

TEST

float, 1 decimal

raw refraction value

16

CONC%

float, selected decimals concentration

value

31

temp C

float, 1 decimal

temperature, degrees centigrade

46

RI 25 C

float, 4 decimals

standard RI at 25 C

61

phase

1 digit

operation phase code

63 error integer

error

code

66

divert

integer

divert control status code

68 checksum

integer

76

terminator

CR and LF