RIGOL Command Systems

Programming Guide for DS1000E, DS1000D Series

2-24

Example:
:TRIG:EDGE:SWE AUTO

Setup the trigger type as AUTO.

:TRIG:EDGE:SWE?

The query returns AUTO.


5. :TRIGger<mode>:COUPling

Command Format:

:TRIGger<mode>:COUPling {DC|AC|HF|LF}
:TRIGger<mode>:COUPling?

Function Explanation:
The commands set and query the coupling type. Thereinto,
DC: Allow all signals pass.
AC: Block DC signals and attenuate the signals lower than 10Hz.
HF: Reject high frequency signals (Higher than 150KHz).
LF: Reject DC signals and attenuate low frequency signals (Lower than 8KHz).
<mode> could be :EDGE, :PULSe or :SLOPe.

Returned Format:
The query returns DC, AC, HF or LF.

Example:
:TRIG:EDGE:COUP DC Setup the coupling type as DC.
:TRIG:EDGE:COUP?

The query returns DC.


6. :TRIGger:HOLDoff

Command Format:
:TRIGger:HOLDoff <count>
:TRIGger:HOLDoff?

Function Explanation:
The commands set and query the trigger holfoff time. The range of <count> is
<count>: 500ns~1.5s.

Returned Format:
The query returns the setting value of <count> in s.