Slope trigger, Slope trigger -32, Trigger:slope:time – RIGOL DS1000D User Manual

Page 44: Trigger:slope:sensitivity

Advertising
background image

RIGOL Command Systems

Programming Guide for DS1000E, DS1000D Series

2-32

SLOPe Trigger

1. :TRIGger:SLOPe:TIME

Command Format:
:TRIGger:SLOPe:TIME <count>
:TRIGger:SLOPe:TIME?

Function Explanation:
The commands set and query the time setting about slope trigger. The range of
<count> is 20ns~10s.

Returned Format:
The query returns the setting value of <count> in s.

Example:
:TRIG:SLOP:TIME 0.01 Setup the slope time is 10ms.
:TRIG:SLOP:TIME?

The query returns 1.000e-02.


2. :TRIGger:SLOPe:SENSitivity

Command Format:
:TRIGger:SLOPe:SENSitivity <count>
:TRIGger:SLOPe:SENSitivity?

Function Explanation:
The commands set and query the trigger sensitive. The range of <count> is:
0.1div ~1div.

Returned Format:
The query returns the setting value of <count> in div.

Example:
:TRIG:SLOP:SENS 0.5 Setup the trigger sensitive as 0.5div
:TRIG:SLOP:SENS? The query returns 5.00e-01