13. :TRIGger:ALTernation:SLOPe:LEVelA

Command Format:
:TRIGger:ALTernation:SLOPe:LEVelA <value>
:TRIGger:ALTernation:SLOPe:LEVelA?

Function Explanation:
The commands set and query the current upper boundary “Level A” of trigger
level in slope trigger mode. The range of <value> is LevelB ~ +6*Scale; Scale
indicates the current vertical scale, the unit is V/div.

Returned Format:
The query returns the setting value of <value> in V.

Example:
:TRIG:ALT:SLOP:LEVA 2 Setup the upper boundary “Level A” of

trigger level as 2V.

:TRIG:ALT:SLOP:LEVA?

The query returns 2.000e+00.


14. :TRIGger:ALTernation:SLOPe:LEVelB

Command Format:
:TRIGger:ALTernation:SLOPe:LEVelB <value>
:TRIGger:ALTernation:SLOPe:LEVelB?

Function Explanation:
The commands set and query the current lower boundary “Level B” of trigger
level in slope trigger mode. The range of <value> is -6*Scale ~ LevelA,Scale
indicates the current vertical scale, the unit is V/div.
Note: The value of Level A (upper boundary) should not be less than the
maximum of Level B (lower boundary).

Returned Format:
The query returns the setting value of <value> in V.

Example:
:TRIG:ALT:SLOP:LEVB -1.5