RIGOL Command Systems

Programming Guide for DS1000E, DS1000D Series

2-74

:DIGital<n>:POSition?

Function Explanation:
The commands set and query the waveform display position of specified channel.
<n> could be 0~15.
When choose Big-waveform, up to 8 waveforms could be displayed on the
screen, and the range of <position> is 0~7;
when choose Small-waveform, up to 16 waveforms could be displayed on the
screen, and the range of <position> is 0~15.

Returned Format:
The query returns the setting value of <position>.

Example:
:DIG2:POS 3

Setup the waveform of digital channel2 displays at the
position of D3.

:DIG2:POS?

The query returns 3.


4. :LA:THReshold

Command Format:
:LA:THReshold<value>[suffix]
:LA:THReshold?

Function Explanation:
The commands set and query the type of threshold of digital channel. <value>
could be TTL, CMOS, ECL or User (User-defined, the range is -8.00~+8.00).
suffix could be V or mV, the minimum interval is 10mV.

Example:
:LA:THR COMS

Setup the type of threshold as COMS.

:LA:THR?

The query returns COMS.


5. :LA:POSition:RESet

Command Format:
:LA:POSition:RESet