Trigger:status, Trig%50, Forcetrig – RIGOL DS1000D User Manual

Page 37

Advertising
background image

Command Systems

RIGOL

Programming Guide for DS1000E, DS1000D Series

2-25

Example:
:TRIG:HOLD 0.0005 Setup the holdoff time as 500ms.
:TRIG:HOLDoff?

The query returns 5.000e-04



7. :TRIGger:STATus?

Command Format:
:TRIGger:STATus?

Function Explanation:
The command queries the operating status of the oscilloscope. The status could
be RUN, STOP, T`D, WAIT or AUTO.

Returned Format:
The query returns RUN, STOP, T'D, WAIT or AUTO.


8. :Trig%50

Command Format:
:Trig%50

Function Explanation:
The command sets the trigger level to the vertical midpoint of amplitude.


9. :FORCetrig

Command Format:
:FORCetrig

Function Explanation:
The command forces the oscilloscope to trigger signal, which is usually used in
“Normal” and “Single” mode.


Advertising
This manual is related to the following products: