Section 3

Page 41

8990startup.fm

Basic Programming Setup

b. Enter the Interval Between Samples. Press

ACCEPT

to continue.

c.

Select either Take First Sample Immediately or After The
First Interval?

Note: When the program is started, the first sample is taken immediately upon

pressing the start button or after the first interval has elapsed.

d. Press

ACCEPT

to continue to Sample Distribution.

Flow Proportional Constant Volume, Variable Time (CVVT)

CVVT sampling is a flow-proportional method of sampling. CVVT samples are
taken when a user-defined flow volume occurs in the flow stream. This occurs
when the sampler is programmed to take a sample every time a specified
volume of flow occurs in the flow stream. Specified flow volume intervals can
vary with the flow rate of the stream, therefore varied time intervals can occur
when constant (fixed) volume occurs.

Flow volume is determined internally, by the optional integral flow meter or by
an external flow meter.

a.

Press

CHANGE CHOICE

until Flow Proportional is displayed. Press

ACCEPT

.

b. In the Flow Proportional menu, press

CHANGE CHOICE

until Constant

Volume, Variable Time is displayed. Press

ACCEPT

.

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

INTERVAL:

CANCEL

00:00 (hrs:min)

CLEAR

ENTRY

ENTER: 000:01—999.00 (hrs:min)

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

TAKE FIRST SAMPLE:

CHANGE

CHOICE

_______________

CANCEL

CHOICES: IMMEDIATE, AFTER 1

st

SAMPLE

11:00 AM 21 - APR - 01

SAMPLE COLLECTION

ACCEPT

CHANGE

UNITS

SAMPLE COLLECTION:

CANCEL

FLOW-PROPORTIONAL

ENTER: 1.00 - 99999999

11:00 AM 21 - APR - 01

FLOW PACING

ACCEPT

CHANGE

CHOICE

FLOW PACING MODE:

CANCEL

CONST VOL / VAR TIME

CHOICES: VAR T / CST VOL, CST T / VAR VOL