Advanced Sampling

18-E. Enter the desired high or low trigger point using the numeric keypad.

Press

ACCEPT

to continue.

18-F. Enter the Deadband value or, if programming for Flow Rate Of Change

or Rainfall, enter a time interval when the flow or rainfall change must
take place (refer to

Alarm Relays Programming on page 91

).

18-G. Enter the number of bottles (out of the total set) to set aside for upset

samples. These will be the last bottles in the tray.

18-H. Select the method of distribution, Sample Per Bottle or

Bottles Per Sample.

18-I. Enter the sample volume using the numeric keypad.

18-J. Press

ACCEPT

to continue to Variable Intervals.