Conductivity Probe

8990options.fm

Section 5

5.5 Conductivity Probe

5.5.1 Conductivity Probe Connection

5.5.2 Conductivity Probe Programming

1.

From the Main Menu, select

OPTIONS>ADVANCED OPTIONS>DATALOG.

2.

Highlight Select Inputs using the

UP

and

DOWN

keys. Press

SELECT

.

3.

Highlight Conductivity (COND.) using the

UP

and

DOWN

. Press

SELECT

.

4.

Press

CHANGE CHOICE

to cycle between Logged and Not Logged, then

press

ACCEPT

.

5.

Enter a logging interval using the numeric keypad, then press

ACCEPT

.

Valid logging intervals are shown on the status bar along the bottom edge
of the display.

6.

Press

CHANGE CHOICE

to select the appropriate units (ms, µs). Press

ACCEPT

to continue.

7.

Select another channel to configure or press

RETURN

to back up one

step. Press

MAIN MENU

to return to the Main Menu display.

5.5.3 Conductivity Temperature Programming

1.

From the Main Menu, select

OPTIONS>ADVANCED OPTIONS>DATALOG.

2.

Highlight Select Inputs using the

UP

and

DOWN

keys and press

SELECT

.

3.

Highlight Conductivity Temperature (COND. TEMP.) using the

UP

and

DOWN

keys, then press

SELECT

.

4.

Press

CHANGE CHOICE

to cycle between Logged and Not Logged, then

press

ACCEPT

.

5.

Enter a logging interval using the numeric keypad, then press

ACCEPT

.

Valid logging intervals are shown on the status bar along the bottom edge
of the display.

6.

Press

CHANGE CHOICE

to select temperature units (°C, °F).

Press

ACCEPT

.

Table 11 Conductivity Probe Wiring (J20)

Pin

Signal Description

Wire Color

A

Probe

Black

B

Probe

Red

C

RTD

White

D

RTD

Green