Step 17 - timed bottle sets – Hach 900 MAX User Manual

Page 59

Advertising
background image

Section 3

Page 57

8990startup.fm

Advanced Sampling

f.

Enter the Program Time Limit. NPDES typically requires monitoring
during the first three hours of any given storm. If the flow volume was
not as high as expected, flow-weighted sampling could continue for
some time as flow rates drop off and sample intervals
become longer.

g. Press

ACCEPT

to continue to Timed Bottle Sets.

Step 17 - Timed Bottle Sets

Timed Bottle Sets enables a single sampler to function like multiple samplers.
Timed Bottle Sets takes a 24-bottle sampler and sample using the first 12
bottles on the first day and the second bottles on the next day.

When a subset of bottles is set aside for each user defined time interval the
sampler will treat that subset as if it were the entire set. At the end of the time
interval the sampler will switch to the next subset and continue sampling.

17-A. Highlight Timed Bottle Sets using the

UP

and

DOWN

keys on the

Advanced Sampling Menu. Press

SELECT

to continue.

17-B. Enable or Disable Timed Bottle Sets using the

CHANGE CHOICE

key.

Press

ACCEPT

to continue.

17-C. Select a method for switching bottle sets. Select either clock time that

switches bottle sets every 24 hours or select duration that sets the
bottles sets in minutes and hours.

The number of bottles that were selected in bottles-per-sample mode are
used as the bottle set size. Two bottles per-sample means two bottles per
Timed Bottle Set.

11:00 AM 21 - APR - 01

SETPOINT SAMPLING

ACCEPT

STORM WATER:

PROGRAM TIME LIMIT:

CANCEL

(hrs:min)

CLEAR

ENTRY

ENTER: 000:01 — 999:00 (hrs:min)

11:00 AM 21 - APR - 01

ADVANCED SAMPLING

SELECT

SPECIAL PUTPUT
START/STOP TIMES
STORM WATER

TIMED BOTTLE SETS

RETURN

UPSET SAMPLE
VARIABLE INTERVALS

11:00 AM 21 - APR - 01

TIMED BOTTLE SETS

ACCEPT

USE CLOCK TIME OR

CHANGE

CHOICE

DURATION:

DURATION (hh:mm)

CANCEL