Step 18 - upset sample – Hach 900 MAX User Manual

Page 60

Advertising
background image

Page 58
Advanced Sampling

8990startup.fm

Section 3

17-D. Enable or Disable Continuous Mode using the

CHANGE CHOICE

key. If

Continuous Mode is enabled the program will continuously run and
switch bottle sets after every specified duration elapses, until manually
stopped. If Continuous Mode is disabled, then the sampling halts when
the last bottle set in the tray is full.

17-E. Press

ACCEPT

to continue to Upset Sample.

Step 18 - Upset Sample

Upset Sampling analyzes the collected samples to determine when samples
meet or exceed specified upset limits. The sampler will swing the distributor
arm around to a sequestered set of bottles and grab an upset bottle.

The upset sample bottles are always the last bottles in the tray. For example, if
there are 24 bottles in the tray, 4 first flush bottles and 4 upset bottles, then the
bottles assignment is 1–4 first flush, 5–20 main program, and 21–24
upset samples.

Note: Unlike Setpoint Sampling,
Upset Sampling can be enabled
while the sampler is performing its
regular sampling program. There
must be more than one bottle in the
sampler to perform Upset Sampling.

Upset Samples are collected even when there are no more regular samples to
be taken; when the status screen says Program Complete, but the bottom line
of the display says Program Running.

Refer to the

Quick Start Guides on page 103

for more setup information for

individual channels.

18-A. Highlight Upset Sampling using the

UP

and

DOWN

keys on the

Advanced Sampling Menu. Press

SELECT

to continue.

18-B. Enable or Disable Upset Sampling using the

CHANGE CHOICE

key.

Press

ACCEPT

to continue.

18-C. Press

CHANGE CHOICE

to select the desired channel to trigger from.

Press

ACCEPT

to make the selection.

18-D. Press either

HIGH CONDITION

or

LOW CONDITION

.

Flow Rate of Change and Rainfall are signals that only increase in
value and do not require Low Condition.

The external control signal (such as an external flow meter) must be
configured in the external equipment for the desired setpoint.

11:00 AM 21 - APR - 01

ADVANCED SAMPLING

SELECT

START/STOP TIMES
STORM WATER
TIMED BOTTLE SETS

UPSET SAMPLING

RETURN

VARIABLE INTERVALS
VARIABLE VOLUME

11:00 AM 21 - APR - 01

UPSET SAMPLES

UPSET CONDITION:

HIGH

CONDITION

LOW

CONDITION