pH Probe

Figure 22

pH Probe Wiring to Junction Box (grounded)

Figure 23

pH Probe Wiring to Junction Box (un-grounded)

5.2.2 pH Probe Programming

1.

From the Main Menu, select

OPTIONS>ADVANCED OPTIONS>DATALOG.

2.

Highlight Select Inputs using the

UP

and

DOWN

keys and press

SELECT

.

3.

Highlight pH using the

UP

and

DOWN

keys, then press

SELECT

.

4.

Press the

CHANGE CHOICE

key to cycle between Logged and Not

Logged, then press

ACCEPT

.

5.

Enter a logging interval, then press

ACCEPT

. Valid logging intervals are

shown on the status bar along the bottom edge of the display.

6.

Select another channel to configure or press

RETURN

to back up one

step. Press

MAIN MENU

to return to the Main Menu display.

5.2.3 pH Probe Calibration

Calibrate the pH probe after the pH probe is connected and programmed.
Calibrating the pH probe requires a thermometer and any two of the following
buffer solutions: 4, 7, or 10 pH.

The pH probe is an application sensitive device. When used in harsh
environments, the accuracy and life expectancy of pH probes can decrease.

Yellow

Green

Clear

pH

Glass

Red

Black

GND

REF

RT
D

RT
D

Yellow

Green

Clear

pH

Glass

Red

GND

REF

RT
D

RT
D