Section 5 optional device installation, 1 rain gauge – Hach 900 MAX User Manual

Page 71

Advertising
background image

Page 69

8990options.fm

Optional Device Installation

Section 5

Optional Device Installation

This section describes how to setup a rain gauge to the Sampler as well as
how to connect, program, calibrate, and maintain the optional water
quality probes:

5.1 Rain Gauge

An external “tipping bucket” rain gauge (Cat. No. 2149) can be connected to
the Rain Gauge connector of the sampler (

Figure 21

). The rain gauge

provides a dry contact closure to the sampler.

As rainfall collects in the 20 cm (8 in.) diameter funnel, it is directed into one
side of a “tipping bucket” assembly. As each bucket fills, it causes the bucket
assembly to tip and empty out into the bottom of the rain gauge. Each tip of
the bucket causes a single contact closure to the rain gauge and sends a
short 12 V dc pulse into pin C of the Rain Gauge connector. Each pulse (tip)
represents 0.025 cm (0.01 in.) of rain.

Figure 21

Rain Gauge Tipping Bucket

Rain Gauge (

section 5.1 on

page 69

Dissolved Oxygen Probe
(

section 5.4 on page 74

)

pH Probe
(

section 5.2 on page 70

)

Conductivity Probe
(

section 5.5 on page 76

)

ORP Probe
(

section 5.3 on page 72

)

Table 7 Rain Gauge Base Board Connections (J5)

Pin

Signal Description

A

+12 V dc source output

B

not used

C

+12 V dc pulse input

D

not used

E

not used

F

not used

Advertising