Advanced Sampling

13-C. Select either Start on Setpoint or Stop on Setpoint by pressing

CHANGE CHOICE

. Press

ACCEPT

to continue.

Start on Setpoint will start a program when the setpoint condition is
met. The program continues to run even if the condition falls back
within the setpoint limits.

Stop On Setpoint halts the program if the setpoint condition falls back
within the setpoint limits and starts again if the limits are exceeded.

13-D. Highlight the desired channel to trigger from, then press

SELECT

.

13-E. Press either

HIGH CONDITION

or

LOW CONDITION

.

Note: Flow Rate of Change and Rainfall are signals that only increase and never

decrease in value, therefore these signals do not require Low Condition.

The external control signal must be configured on the external equipment for
the desired setpoint. An external control device must provide a dry contact
and can include a float switch, push-button, external flow meter, etc. For
interface connections refer to

section 2.10 on page 29

.

13-F. Enable or Disable the trigger point using the

CHANGE CHOICE

key.

13-G. Select the desired high or low trigger point using the numeric keypad.

Press

ACCEPT

. Refer to

Table 3

.

13-H. Enter the Deadband value if required or, if programming Flow Rate Of

Change or Rainfall, enter a time interval that the flow or rainfall change
must take place (refer to

section 6.4 on page 87

).

11:00 AM 21 - APR - 01

SETPOINT SAMPLING

SETPOINT CONDITION:

HIGH

CONDITION

LOW

CONDITION

11:00 AM 21 - APR - 01

SETPOINT SAMPLING

ACCEPT

SETPOINT SAMPLING:

HIGH TRIGGER POINT:

CANCEL

00000 in.

CLEAR

ENTRY

(USE NUMERIC KEYPAD)

11:00 AM 21 - APR - 01

SETPOINT SAMPLING

ACCEPT

SETPOINT SAMPLING

DEADBAND:

CANCEL

0.000

CLEAR

ENTRY

(USE NUMERIC KEYPAD)