Chapter 3 Peak Detection and Labeling

3-22

Applied Biosystems

3

Basic Settings

(spectrum data

only)

Table 3-2 describes the parameters in the Basic Settings tab
of the Spectrum Peak Detection Setup dialog box (see
Figure 3-4 on page 3-14).

Default peak detection values are listed in Section 3.7, Default
Peak Detection Settings
.

Table 3-2 Basic Settings Tab Parameters (Spectrum Data Only)

Parameter

Description

Global Thresholds

%Base Peak
Intensity

NOTE: This
parameter was
previously named
Peak Threshold%
and Base Peak
Relative%.

Specifies a percentage of the base peak intensity as the
threshold value.

To be detected, peaks must be above this threshold and
above the %Max Peak Area value.

If you override Global Thresholds by selecting Use Advanced
Settings (described on page 3-23), the software ignores
Global Thresholds and uses the %Base Peak Intensity,
%Max Peak Area, Minimum Area, and Minimum Intensity
thresholds set on the Advanced tab (described on page 3-28)
to detect peaks for the selected detection range.

NOTE: If you are examining Voyager data that contains a
Low Mass Gate spike, the software may identify the Low
Mass Gate spike as the Base Peak. Truncate the data as
described in Section 5.9, Truncating a Spectrum,
to eliminate
the Low Mass Gate spike and correctly identify the Base
Peak.

Hint: For most Voyager applications, leave this parameter
set at 0 and adjust the %Max Peak Area.

Continued