4 calibration troubleshooting, Calibration troubleshooting – HP Data Explorer 4 Series User Manual

Page 372

Advertising
background image

Chapter 9 Troubleshooting

9-10

Applied Biosystems

9

9.4 Calibration Troubleshooting

Table 9-6 Calibration Troubleshooting—Mariner and Voyager

Symptom

Possible Cause

Action

Auto Calibration is turned
on, but current spectrum
is not auto calibrated

Current spectrum is not
calibrated until the next
time the spectrum is
displayed

Advance to the next
spectrum, then return to
the current spectrum. For
more information, see
Section 5.4, Automatic
Calibration.

During calibration, the
software is not matching
spectrum masses to some
reference masses in the
calibration reference file

The calibration routine
checks peak width to
determine if a peak
matches a Resolved
Isotope Mass or an
Average mass. If narrow
peaks are specified as
Average Masses in the
calibration reference file,
the software mistakes
these narrow peaks as
isotopically resolved and
ignores the reference
mass.

When specifying highly
charged, non-isotopically
resolved species with
peaks less than 1 Da wide
(for example, myoglobin
+20) as reference masses
in a calibration reference
file, set the peak type as a
Resolved Isotope Mass
(even though it is not a
resolved isotope). See
Section 5.3.3, Creating or
Modifying a
Calibration Reference File
(.REF)
.

Mass tolerance or
Minimum Intensity set too
high

Adjust. See Section 5.3,
Manual Calibration.

Advertising